A new reference point for patient dose estimation in neurovascular interventional radiology

2013 
In interventional radiology, dose estimation using the interventional reference point (IRP) is a practical method for obtaining the real-time skin dose of a patient. However, the IRP is defined in terms of adult cardiovascular radiology and is not suitable for dosimetry of the head. In the present study, we defined a new reference point (neuro-IRP) for neuro-interventional procedures. The neuro-IRP was located on the central ray of the X-ray beam, 9 cm from the isocenter, toward the focal spot. To verify whether the neuro-IRP was accurate in dose estimation, we compared calculated doses at the neuro-IRP and actual measured doses at the surface of the head phantom for various directions of the X-ray projection. The resulting calculated doses were fairly consistent with actual measured doses, with the error in this estimation within approximately 15 %. These data suggest that dose estimation using the neuro-IRP for the head is valid.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    19
    References
    3
    Citations
    NaN
    KQI
    []