Preliminary performance characteristics of a dual-energy KCD

1995 
A kinestatic charge detector (KCD) with segmented signal-collection fingers was constructed to evaluate the dual-energy x-ray imaging performance of a KCD. The front segments of the KCD signal-collectors produce a digital low-energy image and the back segments produce a digital high-energy image. A gap between the front and back signal-collectors is used as a filter to increase the separation between the mean energies absorbed in the front and back segments. Preliminary measurements have been performed on the dual-energy KCD to determine its dual-energy imaging characteristics. The KCD output signal has been measured as a function of depth in the chamber. The ion drift velocity, modulation transfer function (MTF), detective quantum efficiency (DQE) and Wiener spectrum have been determined for both the front (low-energy) and back (high-energy) signal detection regions of the KCD.