Calibration of a superconducting gravimeter with an absolute atom gravimeter

2020 
We present in this article a 27 days long common view measurement of an absolute cold atom gravimeter (CAG) and a relative iGrav superconducting gravimeter, which we use to calibrate the iGrav scale factor. We investigate the impact of the duration of the measurement on the uncertainty in the determination of the correlation factor and show that it is limited to about 3\textperthousand~ by the coloured noise of our cold atom gravimeter. A 3 days long measurement session with an additional FG5X absolute gravimeter allows us to directly compare the calibration results obtained with two different absolute meters. Based on our analysis, we expect that with an improvement of its long term stability, the CAG will allow to calibrate the iGrav scale factor to the per mille level after only one day of concurrent measurements.