Myopic Laser Corneal Refractive Surgery Reduces Interdevice Agreement in the Measurement of Anterior Corneal Curvature

2017 
OBJECTIVES: To investigate interdevice differences and agreement in the measurement of anterior corneal curvature obtained by different technologies after laser corneal refractive surgery. METHODS: The prospective study comprised 109 eyes of 109 consecutive patients who had undergone laser-assisted in situ keratomileusis (LASIK). Preoperative and postoperative corneal parameters were measured by Scheimpflug imaging (Pentacam), Placido-slit-scanning (Orbscan) and auto-keratometry (IOLMaster). Preoperative and postoperative anterior corneal curvatures (K readings) were compared between devices. Interdevice agreement was evaluated by Bland-Altman analysis. RESULTS: Preoperatively, the difference of K reading for Pentacam-IOLMaster (0.04±0.20 D) was not statistically significant (P=0.059). The differences between Pentacam-Orbscan and Orbscan-IOLMaster were 0.20±0.34 D (P<0.001) and -0.17±0.29 D (P<0.001), respectively. After surgery, no difference was found for Pentacam-Orbscan (-0.05±0.38, P=0.136). The differences between Pentacam-IOLMaster and Orbscan-IOLMaster were 0.13±0.29 D (P<0.001) and 0.19±0.34 D (P<0.001). Preoperative interdevice agreement (95% limit of agreement [LOA]) between Pentacam and Orbscan, Pentacam and IOLMaster, and Orbscan and IOLMaster were 1.31 D, 0.79 D and 1.14 D, respectively. The 95% LOAs decreased to 1.47 D, 1.14 D, and 1.34 D after refractive surgery. CONCLUSION: Corneal refractive surgery changed the preoperative and postoperative interdevice differences in corneal curvature measurements and reduced interdevice agreement, indicating that the devices are not interchangeable.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    35
    References
    2
    Citations
    NaN
    KQI
    []