A reliable aptamer array prepared by repeating inkjet-spotting toward on-site measurement

2016 
Abstract A preparation protocol is proposed for a reliable aptamer array utilizing an ink-jet spotter. We accumulated streptavidin and biotinylated-aptamer in this order on a biotinylated-polyethylene glycol-coated gold substrate to prepare an aptamer array. The aptamer array was prepared with an alternate spotting structure where each aptamer spot was placed between reference spots formed with blocking solution thus suppressing contamination from neighboring spots during the blocking and washing processes. Four aptamer spots were prepared in a small area of 1×4.8 mm 2 with five reference spots made of blocking solution. We evaluated the thrombin binding ability of the spotted aptamer array using a multi-analysis surface plasmon resonance sensor. We prepared a disposable capillary-driven flow chip designed for on-site measurement ( Miura et al., 2010 ) with our aptamer array and detected thrombin from phosphate-buffered saline at concentrations of 50 ng mL −1 and 1 μg mL −1 (equivalent to 1.35 and 27 nM, respectively). A correlation was observed between the refractive index shift and thrombin concentration. This implies that our array preparation protocol meets the requirement for the preparation of a one-time-use chip for on-site measurement.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    50
    References
    6
    Citations
    NaN
    KQI
    []