The adjustable systemic-pulmonary artery shunt provides precise control of flow in vivo.

2010 
The ratio of pulmonary:systemic blood flow (Qp:Qs) remains problematic after single ventricle reconstruction. The adjustable systemic-pulmonary artery shunt (AS) was created as a solution for this problem. Prototype ASs were created using a screw-plunger mechanism as a variable resistor. A stepper motor controls plunger displacement. Six adult dogs underwent placement of a 4-mm AS in the femoral position to test its ability to control flow. Shunts were placed as arteriovenous fistulae to simulate the continuous flow of systemic-pulmonary AS. The 3.5-mm control shunts (CS) were placed on the contralateral side. The stepper motor was rotated from fully open to 3.4 mm of plunger depression for six complete cycles. Flow in the fully open AS was 687.9 ± 28.7 cc/min* vs. 578.7 ± 26.8 cc/min in the CS (flow ± standard error, *p