Difference between two exhaled nitric oxide analyzers, NIOX VERO® electrochemical hand-held analyzer and NOA280i® chemiluminescence stationary analyzer

2019 
AbstractBackground: Fractional exhaled nitric oxide (FENO) is useful for the evaluation of eosinophilic airway inflammation, including that seen in asthma. Although a new electrochemical hand-held FENO analyzer, the NIOX VERO® (Aerocrine AB, Solna, Sweden), is clinically convenient to use, it has not been fully compared with the chemiluminescence stationary electrochemical analyzer NOA280i® (Sievers Instruments, Boulder, CO, USA) in terms of the level of measured FENO. The aim of this study was to determine whether there is a difference between the two analyzers. Methods: The FENO levels measured with both NIOX VERO® and NOA280i® were evaluated in 1,369 adults at Juntendo University Hospital from May 2016 to October 2016. Results: The median FENO level measured with the NIOX VERO® was significantly lower than that measured with the NOA280i® (41 ppb, range 5–368 ppb vs. 29 ppb, range 5–251 ppb; p < 0.001). There was a strong positive correlation in the measurement of FENO level between the NOA280i® and the...