Investigation of the Lag Effect in X-Ray Flat-Panel Detector for Cone-Beam Computed Tomography

2020 
The purpose of this work is to investigate the factors determining the magnitude and nature of the lag effect (residual signal) in the Varian PaxScan 4343CB flat-panel detector. Depending on the detector mode, typical value of residual signal is 1-2% after single exposed frame, increasing up to 3% after multiple exposures. The lag effect is the cause of up to 3.6% error of average flat-field signal measurement if used for averaging series of continuously exposed frames, which leads to incorrect assessment of X-ray attenuation coefficients. Spatial dependence of the lag effect is confirmed; the results of correction of projection images using spatially localized impulse response functions are demonstrated.