Transient out-of-range impedance in "hybrid" implantable cardioverter-defibrilator (ICD) system: a case series.

2021 
A retrospective analysis of 60 patients with hybrid ICD systems: Boston Scientific device paired with non-Boston leads. In 10 (17%) patients transient, out-of-range peaks of ventricular pace impedance trend were observed. Probable cause is header-lead interaction incompatibility. This matter is known mainly for pacemakers systems but not for ICDs. Investigation this issue is crucial because consequences in ICD systems are unpredictable and risk might be higher than in pacing systems. This article is protected by copyright. All rights reserved.