Time‐optimized 4D phase contrast MRI with real‐time convex optimization of gradient waveforms and fast excitation methods

2019 
PURPOSE: To shorten 4D flow acquisitions by shortening TRs with fast RF pulses and gradient waveforms. Real-time convex optimization is used to generate these gradients waveforms on the scanner. THEORY AND METHODS: RF and slab-select waveforms were shortened with a minimum phase SLR excitation and the time-optimal variable-rate selective excitation method. Real-time convex optimization was used to shorten bipolar and spoiler gradients by finding the shortest gradient waveforms that satisfied constraints on scan parameters, gradient hardware, M0 , M1 , and peripheral nerve stimulation. Waveforms were calculated and TE and/or TR values were compared for a range of scan parameters and compared to a conventional 4D flow sequence. The method was tested in flow phantoms, and in the aorta and neurovasculature of volunteers (N = 10). Additionally, eddy current error was measured in a large phantom. RESULTS: TEs and TRs were shortened by 21-32% and 20-34%, respectively, compared to the conventional sequence over a range of scan parameters. Bland-Altman analysis of 2 flow phantom configurations showed flow rate bias of 0.3 mL/s and limits of agreement (LOA) of [-6.9, 7.5] mL/s for a cardiac phantom and a bias of -0.1 mL/s with LOA = [-0.4, 0.2] mL/s for a neuro phantom. Similar agreement was also seen for flow measurements in volunteers (bias = -1.0 and -0.1 mL/s, LOA = [-34.9, 33.0] and [-0.7, 0.6] mL/s). Measured eddy currents were 39% larger with the CVX + mpVERSE method. CONCLUSION: The real-time optimized 4D flow gradients and fast slab-selection excitation methods produced up to 34% faster TRs with excellent flow measurement agreement compared to a conventional 4D flow sequence.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    30
    References
    9
    Citations
    NaN
    KQI
    []