Evaluation of a continuously active disinfectant for decontamination of portable medical equipment.
2021
A single spray application of a continuously active disinfectant on portable equipment resulted in significant reductions in aerobic colony counts over 7 days and in recovery of Staphylococcus aureus and enterococci: 3 of 93 cultures (3%) versus 11 of 97 (11%) and 20 of 97 (21%) in quaternary ammonium disinfectant and untreated control groups, respectively.