Anatomical Joint Measurement with Application to Medical Robotics

2020 
Robotic-assisted orthopaedic procedures demand accurate spatial joint measurements. Tracking of human joint motion is challenging in many applications, such as in sport motion analyses. In orthopaedic surgery, these challenges are even more prevalent, where small errors may cause iatrogenic damage in patients – highlighting the need for robust and precise joint and instrument tracking methods. In this study, we present a novel kinematic modelling approach to track any anatomical points on the femur and / or tibia by exploiting optical tracking measurements combined with a priori computed tomography information. The framework supports simultaneous tracking of anatomical positions, from which we calculate the pose of the leg (joint angles and translations of both the hip and knee joints) and of each of the surgical instruments. Experimental validation on cadaveric data shows that our method is capable of measuring these anatomical regions with sub-millimetre accuracy, with a maximum joint angle uncertainty of ±0.47°. This study is a fundamental step in robotic orthopaedic research, which can be used as a ground-truth for future research such as automating leg manipulation in orthopaedic procedures.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    31
    References
    5
    Citations
    NaN
    KQI
    []