Determination of dioxin-like polychlorinated biphenyls in 1 mL whole blood using programmable temperature vaporization large volume injection coupled to gas chromatogram and high-resolution mass spectrometry.

2013 
Abstract A sensitive method based on programmable temperature vaporization large volume injection coupled to gas chromatogram and high-resolution mass spectrometry (PTV-GC–HRMS) has been developed for the determination of ultra trace levels of dioxin-like polychlorinated biphenyls (DL PCBs) in small amounts of human blood. Blood samples (1 mL) were first extracted by column extraction and then purified with column chromatorgraphies. Final extracts (20 μL) were introduced to the PTV injector under the solvent vent mode and detected by GC–HRMS (SIM mode). PTV parameters were observed by changing one factor at a time (practical conditions: vent flow: 50 mL min −1 , vent pressure: 0 kPa and vent time: 0.1 min), recoveries of most PCB congeners ranged from 55.1% to 108%, and method detection limits were in the range of 0.11–1.63 pg g −1 .