THE EVALUATION OF THE ESTABLISHED SAFE ZONE IN PATIENTS WITH ARTICULAR SURFACE REPLACEMENT PROSTHESIS

2018 
IntroductionStudies of metal-on-metal (MoM) hip resurfacing arthroplasty (HRA) have reported high complication and failure rates due to elevated metal ion levels. These rates were shown to be especially high for the Articular Surface Replacement (ASR) HRA, possibly due to its unique design. Associations between metal ion concentrations and various biological and mechanical factors have been reported. Component positioning as measured by acetabular inclination has been shown to be of especially strong consequence in metal ion production in ASR HRA patients, but few studies have evaluated acetabular anteversion as an independent variable.The primary objective of this study was to evaluate the associations between component orientation, quantified by acetabular inclination and anteversion, and blood metal ions. Secondly, we sought to report whether conventional safe zones apply to MoM HRA implants or if these implants require their own positioning standards.MethodsWe conducted a multi-center, prospective stu...