Comparison of porous titanium-surfaced and standard smooth-surfaced bone plates and screws in an unstable fracture model in dogs.

1986 
: Bilateral midshaft femoral osteotomies were fixed with a 3-mm fracture gap in 6 dogs. In each dog, one femur was fixed with a porous titanium-surfaced bone plate and screws, whereas the opposite femur was fixed with a standard smooth-surfaced bone plate and screws. The mean removal torque for porous titanium-surfaced screws (32.3 kg X cm) was significantly (P less than 0.01) greater than the mean removal torque for standard screws (4.4 kg X cm). Osseous tissue ingrowth into the surface of porous titanium-surfaced screws was verified by histologic examination of the bone-screw interface. Radiographic and histologic examinations of the osteotomy gaps showed accelerated primary gap healing in osteotomies fixed with porous titanium-coated implants, compared with slower callus healing seen in osteotomies fixed with standard smooth-surfaced implants.