Repeatability and intermediate precision of a mass concentration calibration system
2019
AbstractMany aerosol instruments require calibration to make accurate measurements. A centrifugal particle mass analyzer (CPMA) and aerosol electrometer can be used to calibrate aerosol instruments that measure mass concentration. To understand the sources of uncertainty in the calibration method, two CPMA-electrometer systems were tested to measure the repeatability and intermediate precision of the system, where the repeatability is the standard deviation of several measurements using the same system over a short period of time, and the intermediate precision is the standard deviation of several measurements using different instruments with different calibrations over a long period of time. It was found that the repeatability of the CPMA and the aerosol electrometer were both 0.8%, while the intermediate precision was 1.3% and 2.2%, respectively. The intermediate precision of the aerosol electrometers determined here compares well with a broader study by national metrology institutes which determined an...