Construction and Calibration of a Streaked Optical Spectrometer for Shock Temperature.

2020 
Here we describe the implementation and calibration of a streaked visible spectrometer (SVS) for optical pyrometry and emission/absorption spectroscopy on light gas gun platforms in the UC Davis Shock Compression Laboratory. The diagnostic consists of an optical streak camera coupled to a spectrometer to provide temporally and spectrally-resolved records of visible emission from dynamically-compressed materials. Fiber optic coupling to the sample enables a small diagnostic footprint on the target face and flexibility of operation on multiple launch systems without the need for open optics. We present the details of calibration (time, wavelength and spectral radiance) for absolute temperature determination and present benchmark measurements of system performance.