The rate of sural nerve violation using the Achillon device: a cadaveric study.

2013 
Background:Management of acute Achilles tendon rupture remains controversial. Open repair results in lower rerupture rates but is associated with complications from wound healing. Percutaneous and limited open repairs minimize soft tissue complications; however, there is a theoretical increased risk of sural nerve injury. We evaluated the risk of sural nerve violation with the Achillon limited open repair device in cadavers.Methods:This cadaveric study consisted of 2 parts. The first part evaluated the frequency and location of sural nerve violation using the standard Achillon protocol in 18 intact embalmed left lower limbs. The second part involved device manipulation with 30 degrees of either internal or external rotation in 13 paired, intact embalmed lower limbs.Results:In the first part, 5 of 18 specimens had at least 1 suture passing through the sural nerve. Eight of the 54 needle passes (14.8%) directly pierced the substance of the sural nerve. With internal rotation, 8 of 13 specimens had at least ...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    37
    References
    25
    Citations
    NaN
    KQI
    []