TBC: A simple algorithm to rule out abnormalities in electrocardiograms of patients with pacemakers

2013 
Background: The aim of the study was to create a straightforward method to rule out abnormalities in electrocardiograms (ECGs) performed in patients with pacemakers. Methods: The TBC method screens the ECG for any of the following findings: Tachycardia with pacing spikes, Bradycardia without spikes and Chaos with spikes unrelated to QRS-T complexes. T was considered to advise for patient assessment and B and C to require referral for urgent pacemaker evaluation. The diagnostic accuracy of the algorithm was validated using a cohort of 151 ECGs with normal and dysfunctional pacemakers. The effect of the algorithm was then evaluated  for diagnostic skills and management of patients with pacemakers by non-cardiologists, comparing their diagnostic accuracy before and after teaching the algorithm. Results: The TBC algorithm had a sensitivity of 86% and a specificity of 94% in diagnosing a malfunctioning pacemaker. The diagnostic skills and patient referral were significantly improved (74.8% vs. 89.5%, p < 0.001; and 57.4% vs. 83%, p < 0.001). Conclusions: TBC is an easy to remember and apply method to rule out severe abnormalities in ECGs of patients with pacemakers. TBC algorithm has a very good diagnostic capability and is easily applied by non-expert physicians with good results.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    0
    References
    1
    Citations
    NaN