Development of a Prototype Detector Using APD-Arrays Coupled With Pixelized Ce:GAGG Scintillator for High Resolution Radiation Imaging

2014 
A novel digital PET scanner based on Time over Threshold method is developed. The positron emission tomography (PET) is composed of 144channel Ce:Gd 3 Al 2 Ga 3 O 12 (GAGG)-Avaranche photodiode (APD) detector arrays individually coupled with custom designed Time over Threshold (ToT) application-specific integrated circuit (ASIC) to realize the high count rate and good spatial resolution. Such an imaging system provides a simple front-end circuit and flexible digital signal processing like multiplexing such as a pulse train method. The measured energy resolution of the detector system was 6.7% for the 511 keV peak, and 4.25 ns time resolution was measured with a single detector module. The measured spatial resolution for a point source was 1.37 mm FWHM for our initial data with a columnar 22 Na source.