Late distortion of the original Palmaz stent implanted in postoperative lesions associated with congenital heart disease

2005 
The objective of this study was to report late distortion of a Palmaz stent. Late distortion of an original Palmaz stent, implanted in an extracardiac lesion, is rare. We completed a 1-year follow-up of 54 patients who had been implanted with 80 Palmaz stents in extracardiac lesions. Distortion of two stents was detected in two patients. For case 1, we implanted a P188 stent for supravalvar pulmonary stenosis complicating an arterial switch operation in a 14-year-old girl. Seven months later, we found compression of the stent. Although we implanted two P308 stents anterior to the distorted stent, distortion of both stents developed after 1 month. Two more P308 stents placed inside each stent were gradually recompressed. A CAT scan showed compression of the stent by a dilated sinus of valsalva. For case 2, we implanted a P308 stent for stenosis of the superior vena cava after Williams operation in an 11-year-old boy. A chest X-ray documented longitudinal compression of the stent 27 months after implantation and a CAT scan showed the ascending aorta was in contact with the stent. A Palmaz stent may be distorted when implanted in a lesion adjacent to a pulsating aorta. © 2005 Wiley-Liss, Inc.