Presumptive Tricuspid Valve Malfunction Induced by a Pacemaker Lead: A Case Report and Review of the Literature
1980
A 23-year-old woman developed 3° AV block with syncope. Insertion of a permanent pacemaker lead was followed by the onset of a persistent murmur in late systole preceded by single or multiple clicks. The murmur was best heard at the left sternal edge, grade 3–4/6 with two major frequencies (60–250 Hz), increased with inspiration and on assuming the erect posture. It was considered to be tricuspid in origin and related to interference of the tricuspid valve apparatus by the pacemaker lead resulting in tricuspid regurgitation. No tricuspid valve prolapse or flutter was seen on echocardiography. Withdrawal of the pacemaker lead resulted in immediate disappearance of the new auscultatory findings. Review of the literature suggests that the appearance of such a murmur following pacemaker insertion could be associated with later complications in relation to tricuspid valve dysfunction. It is therefore recommended that, under these circumstances, permanent pacemaker leads should be appropriately repositioned.