Optimization and quality control of computed radiography

1994 
Computed radiography (CR) is a relatively new technique for projection radiography. Few hospitals have CR devices in routine service and only a handful have more than one CR unit. As such, the clinical knowledge base does not yet exist to establish quality control (QC) procedures for CR devices. Without assurance that CR systems are operating within nominal limits, efforts to optimize CR performance are limited in value. A complete CR system includes detector plates that vary in response, cassettes, an electro-optical system for developing the image, computer algorithms for processing the raw image, and a hard copy output device. All of these subsystems are subject to variations in performance that can degrade image quality. Using CR manufacturer documentation, we have defined acceptance protocols for two different Fuji CR devices, the FCR 7000 and the AC1+, and have applied these tests to ten individual machines. We have begun to establish baseline performance measures and to determine measurement frequencies. CR QC is only one component of the overall quality control for totally digital radiology departments.© (1994) COPYRIGHT SPIE--The International Society for Optical Engineering. Downloading of the abstract is permitted for personal use only.