Alignment and calibration of the ICON-FUV instrument: development of a vacuum UV facility

2016 
The optical calibration of the ICON-FUV instrument requires designing specific ground support equipment (GSE). The ICON-FUV instrument is a spectrographic imager that operates on two specific wavelengths in the UV (135.6 nm and 157 nm). All the operations have to be performed under vacuum UV light. The optical setup is based on a VUV monochromator coupled with a collimator that illuminates the FUV entrance slit. The instrument is placed on a manipulator providing fields pointing. Image quality and spectral properties can be then characterized for each field. OGSE, MGSE, optical calibration plan and vacuum alignment of the instrument are described.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    6
    References
    1
    Citations
    NaN
    KQI
    []