Microfluidic method of pig oocyte quality assessment in relation to different follicular size based on lab-on-chip technology.

2014 
>5 mm), medium (3–5 mm), and small (<3 mm), were analyzed after preceding in vitro maturation (IVM). The LOC analysis was performed using a silicon-glass sandwich wi th two glass optical fibers positioned “face-to-face.” Oocytes collected from follicles of different size classes revealed specific and distinguishable spectral characteristics. The absorbance spectra (microspectrometric specificity) for oocytes isolated from large, medium, and small follicles differ significantly ( �� < 0.05 )a nd the absorbance wavelengths were between 626 and 628 nm, between 618 and 620 nm, and less than 618 nm, respectively. The present study offers a parametric and objective method of porcine oocyte assessment. However, up to now this study has been used to evidence spectral markers associated with follicular size in pigs, onl y. Further investigations with functional-biological assays and comparing LOC analyses with fertilization and pregnancy success and the outcome of healthy offspring must be performed.