Development and validation of a competitive immunoassay for urinary S-phenylmercapturic acid and its application in benzene biological monitoring

2002 
An immunoassay that quantifies urinary S-phenylmercapturic acid (PMA), a benzene-specific biomarker, has been developed and its potential usefulness as a screening tool for monitoring occupational exposure to benzene has been demonstrated. Analytical reliability has been confirmed by correlation of results with gas chromatography-mass spectrometry (GC/MS) data (R = 0.92). The assay has been configured as a competitive enzyme-linked immunosorbent assay (ELISA) to facilitate rapid throughput of samples. The ELISA has a working range of 40-1200 nmol l-1 urinary PMA and appears to be unaffected by the presence of structurally related urinary metabolites. Background levels of 0-1.9μmol PMA/mol creatinine (mean 0.9 μmol mol-1, n = 32) were measured in non-smoking control subjects. Recent exposures to benzene (8 h time-weighted averages-TWA), during diverse industrial processes, over the range 0-4.8ppm were identified by application of the assay in biological monitoring programmes.