LC–MS/MS assay for the quantitation of the ribonucleotide reductase inhibitor triapine in human plasma

2017 
Abstract The ribonucleotide reductase inhibitor and radiosensitizer triapine (3-aminopyridine-2-carboxaldehyde thiosemicarbazone (3-AP), NSC 663249) is clinically being evaluated via the intravenous (IV) route for the treatment of cervical and vulvar cancer in combination with primary cisplatin chemoradiation. The need for a 2-h infusion and frequent administration of triapine is logistically challenging, prompting us to pursue oral (PO) administration. In support of the clinical trial investigating oral triapine in combination with chemoradiation, we developed and validated a novel LC–MS/MS assay for the quantification of triapine in 50 μL human plasma. After protein precipitation, chromatographic separation of the supernatant was achieved with a Shodex ODP2 column and an isocratic acetonitrile-water mobile phase with 10% ammonium acetate. Detection with an ABI 4000 mass spectrometer utilized electrospray positive mode ionization. The assay was linear from 3 to 3,000 ng/mL and proved to be accurate (97.1–103.1%) and precise (
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    19
    References
    3
    Citations
    NaN
    KQI
    []