Automated Tube Voltage Selection for Radiation Dose Reduction in CT Angiography Using Different Contrast Media Concentrations and a Constant Iodine Delivery Rate

2015 
OBJECTIVE. The purpose of this study was to systematically investigate radiation dose reduction using automated tube voltage selection during CT angiography (CTA) and to evaluate the impact of contrast medium (CM) injection protocols on dose reduction. MATERIALS AND METHODS. A circulation phantom containing the thoracic and abdominal vasculature was used. Four different concentrations of CM (iopromide 300 and 370 mg I/mL and iomeprol 350 and 400 mg I/mL) were administered while maintaining an identical iodine delivery rate (1.8 g I/s) and total iodine load (20.0 g). Three different scanning protocols for CTA of the thoracoabdominal aorta were used: protocol A, no dose modulation; protocol B, automated tube current modulation (CARE Dose4D); and protocol C, automated tube voltage selection (CARE kV). The dose-length product was recorded to calculate the effective dose. Attenuation values (in Hounsfield units), image noise levels, and signal-to-noise ratios (SNRs) in six predefined intravascular sites (three...