Early experience with the new endovascular aneurysm sealing system Nellix: First clinical results after 50 implantations:

2016 
ObjectiveIn the present study, 50 EVAS procedures were evaluated in regard to primary (survival and technical success) and secondary (device-related complications) events of interest.MethodsThe single center study was conducted from July 2013 to August 2014 with prospective collection of the clinical data. The clinical results were controlled by CT angiography and contrast-enhanced ultrasound.ResultsThe technical success was 98% and the 30-day mortality 4%. One (2%) patient died from multisystem organ failure and another patient from an intracranial bleeding, respectively. One patient (2%) suffered from a device-related aneurysm rupture. During early follow-up, one (2%) patient developed an endoleak type II, while three (6%) patients suffered from a partial endograft limb thrombosis. Overall, a secondary intervention was necessary in six (12%) patients.ConclusionsWith the Nellix EVAS system, a high primary technical success of 98% was achieved; one (2%) patient developed an endoleak type II which did not ...