A simple apparatus for safety assessment of magnetically induced torque on active implantable medical devices (AIMDs) under 1.5 T and 3.0 T MRI.
2021
OBJECTIVE Our goal is to design and validate a simple apparatus for the safety assessments of magnetically induced torques by four active implantable medical devices (AIMDs) in 1.5 T and 3.0 T magnetic resonance imaging (MRI) environments. MATERIALS AND METHODS A simple apparatus was designed to measure the magnetically induced torque. Before testing the apparatus, its effectiveness was first evaluated with a commercially available weights. The torque values of the four AIMDs, namely implantable cardioverter defibrillator (ICD), pacemaker, bone conduction implant, and cochlear implant, under 1.5 T and 3.0 T MRI environments were measured and compared with established international standard. RESULTS The ICD, pacemaker, bone conduction implant, and cochlear implant had average torques of 1.405, 0.255, 9.460, and 5.490 N·cm under 1.5 T MRI and 2.950, 0.668, 14.618, and 11.484 N·cm under 3.0 T MRI, respectively. DISCUSSION The apparatus developed in this study was verified to be feasible for measuring magnetically induced torques, with tests on four types of AIMDs under 1.5 T and 3.0 T MRI environments. Although the designed structure is simple, the apparatus was shown to measure torque values of AIMDs under MRI accurately.