Comparison of the triple-point temperatures of {sup 20}Ne, {sup 22}Ne and normal Ne

2013 
At the National Metrology Institute of Japan (NMIJ), the triple points of 20Ne and 22Ne were realized using modular sealed cells, Ec3Ne20 and Ec8Ne22, made by the Istituto Nazionale di Ricerca Metrologica (INRiM) in Italy. The difference of the triple-point temperatures of 20Ne and 22Ne was estimated by using the sub-range of standard platinum resistance thermometers (SPRTs) calibrated by NMIJ on the International Temperature Scale of 1990 (ITS-90). The melting curves obtained with the Ec3Ne20 and Ec8Ne22 cells show narrow widths (0.1 mK) over a wide range of the inverse of the melted fraction (1/F) from 1/F=1 to 1/F=10. The liquidus point Ttp estimated by the melting curves from F∼0.5 to F∼0.85 using the Ec8Ne22 is 0.146 29 (4) K higher than that using the Ec3Ne20 cell, which is in good agreement with that observed by INRiM using the same cells. After correction of the effect of impurities and other isotopes for Ec3Ne20 and Ec8Ne22 cells, the difference of Ttp between pure 20Ne and pure 22Ne is estimated...