A small-size transfer blackbody cavity for calibration of infrared ear thermometers

2014 
Abstract A small-size transfer blackbody cavity for calibration of infrared ear thermometers (IRETs) was developed and characterized at the Korea Research Institute of Standards and Science. This blackbody cavity consists of a reflector exposed to the air and a radiator with three-step curves immersed in a water-bath, and has an angularly uniform emissivity of higher than 0.9993. The radiance temperature of the blackbody cavity was measured with an IRET. We also calculated the effective emissivity by using the software STEEP322, considering the influence of the shape and temperature of the probe-tip of the IRET on the effective emissivity of the blackbody cavity. The measured and calculated radiance temperatures of the blackbody cavity were compared to those of the ASTM-type blackbody cavity and are in good agreement. Uncertainties (k = 1) of the blackbody cavity are estimated to be less than 44 mK in the temperature range 35-42 °C.