Miniaturized force-indentation depth sensor for tissue abnormality identification during laparoscopic surgery

2010 
This paper presents a novel miniaturized force-indentation depth (FID) sensor designed to conduct indentation on soft tissue during minimally invasive surgery. It can intra-operatively aid the surgeon to rapidly identify the tissue abnormalities within the tissue. The FID sensor can measure the indentation depth of a semi-spherical indenter and the tissue reaction force simultaneously. It make use of with fiber optical fiber sensing method measure indentation depth and force and is small enough to fit through a standard trocar port with a diameter of 11 mm. The created FID sensor was calibrated and tested on silicone block simulating soft tissue. The results show that the sensor can measure the indentation depth accurately and also the orientation of the sensor with respect to the tissue surface whilst performing indentation.