Compression forces of internal and external ankle fixation devices with simulated bone resorption.

2010 
Background: Internal and external fixation techniques have been developed to provide rigidity and stability to a fusion site such as in tibiotalocalcaneal (TTC) arthrodesis. Compression of the fusion site plays an integral role in primary bone healing, though little work has been done to quantify the compressive force values of ankle fixation devices. Materials and Methods: Using synthetic and cadaveric bone models, a Newdeal/Integra PantaNail and DePuy VersaNail were tested as compressive intramedullary (IM) nails while an Encore True/Lok and an Ace-Fischer frame were tested as external fixators. Results: The PantaNail experienced maximum compressive loads of 1898 and 1255 N in synthetic and cadaveric constructs, respectively. The VersaNail experienced max compressive loads 388 N during installation. All IM nails tested experienced decreased compressive loads after removal of the external guide and instrumentation. The external fixators were loaded to approximately 1200 N in both synthetic and cadaveric ...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    29
    References
    39
    Citations
    NaN
    KQI
    []