The use of ultra-low-dose CT scans for the evaluation of limb fractures: is the reduced effective dose using ct in orthopaedic injury (REDUCTION) protocol effective?

2016 
AimsTo evaluate whether an ultra-low-dose CT protocol can diagnose selected limb fractures as well as conventional CT (C-CT).Patients and MethodsWe prospectively studied 40 consecutive patients with a limb fracture in whom a CT scan was indicated. These were scanned using an ultra-low-dose CT Reduced Effective Dose Using Computed Tomography In Orthopaedic Injury (REDUCTION) protocol. Studies from 16 selected cases were compared with 16 C-CT scans matched for age, gender and type of fracture. Studies were assessed for diagnosis and image quality. Descriptive and reliability statistics were calculated. The total effective radiation dose for each scanned site was compared.ResultsThe mean estimated effective dose (ED) for the REDUCTION protocol was 0.03 milliSieverts (mSv) and 0.43 mSv (p < 0.005) for C-CT. The sensitivity (Sn), specificity (Sp), positive predictive value (PPV) and negative predictive value (NPV) of the REDUCTION protocol to detect fractures were 0.98, 0.89, 0.98 and 0.89 respectively when tw...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    33
    References
    15
    Citations
    NaN
    KQI
    []