Evaluation des performances du cytomètre « MUSE AUTO CD4/CD4% » vs « GUAVA AUTO CD4/CD4% » pour la mesure du taux de lymphocytes CD4 chez des patients infectés par le VIH au Cameroun

2019 
Introduction: This study aimed to evaluate the performances of the MUSE® flow cytometer compared with the reference GUAVA® flow cytometer. Methods: We conducted an experimental study on HIV-infected patient samples. Venous blood samples, collected in a K3 EDTA tube, were analyzed within 24-48 hours by MUSE® and GUAVA® cytometers at the International Center for medical diagnosis (Centre International de Diagnostic medical) in Yaounde. Results: In total, 227 samples were analyzed. There was a strong intraclass correlation (p<0.0001) between MUSE® and GUAVA® cytometers with a correlation coefficient 0.998 (95% CI: 0,998-0,999) for the absolute values and 0,992 (95% CI: 0,989-0,994) for the percentages. A strong positive linear correlation (p=0.0001) was found between MUSE® and GUAVA® cytometers with linear regression slope r2 = 0.98 (95% CI=0,97-0,99) for the absolute values and r2= 0.98 (95% CI= 0.96-1,00) for the percentages. The biases were -4,80 cells/µl (-101.31-91.71) for the absolute values and -0.89% (IC: -6,08-4.3) for the percentages. The percentage of data points outside the limits of agreement was 12/227 (5.29%) and 10/227 (4.41%) respectively for the absolute values and percentages. Cohen's kappa coefficient was 0.92 and the area under the curve was 0,9975 (CI 95%: 0.99-1). Conclusion: MUSE®AUTO CD4/CD4% cytometer is a powerful instrument because its results are consistent with those obtained by the reference cytometer. It can enable tracking of patients infected with HIV, in particular in the developing countries.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    0
    References
    0
    Citations
    NaN
    KQI
    []