High-voltage capillary zone electrophoresis of red blood cells

1989 
Abstract The high-voltage wide-bore capillary zone electrophoresis of red blood cells was investigated. The reproducibility of the retention time (electrophoretic mobility) is excellent and the differentiation among various species is good. The peaks in the electropherogram describe the distribution of the size and/or surface charge of the cells and are therefore broad. The relationship between the peak height and the number of cells injected is good, with linear correlation coefficients better than 0.98. Details of the preparation of cell suspensions and support electrodes are given, which is essential for obtaining reproducible results. The inner surface of FEP capillary tubing is degraded by the application of high voltage and a pause is necessary between successive experiments if good and reproducible peak shapes are to be obtained. The length of the pause increases with the number of experiments made, and finally the tubing becomes useless. Inspection of the inner surface by X-ray photoelectron spectroscopy showed the breakdown of CHF bonds, but the actual mechanism is not known.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    8
    References
    28
    Citations
    NaN
    KQI
    []