Biomechanical Parameters of Mesh Reinforcement and Analysis of a Novel Device for Incisional Hernia Prevention.

2021 
BACKGROUND Prophylactic mesh augmentation (PMA) is an effective technique utilized to reduce the risk of incisional hernia. This study analyzes the biomechanical characteristics of a mesh-reinforced closure and evaluates a novel prophylactic mesh implantation device (SafeClose Roller System; SRS). MATERIALS AND METHODS A total of eight senior-level general surgery trainees (≥4 years of training) from the University of Pennsylvania Health System participated in the study. Biomechanical strength, mesh stiffness, mesh uniformity, and time efficiency for fixation were compared among hand-sewn mesh fixation, SRS mesh fixation and a no-mesh fixation control. Porcine abdominal wall specimens served as simulated laparotomy models. RESULTS Biomechanical load strength was significantly higher for mesh reinforced repairs (P = 0.009). The SRS resulted in a stronger biomechanical force than hand-sewn mesh (21.2 N stronger, P = 0.317), with more uniform mesh placement (P < 0.01), faster time of fixation (P < 0.001) and with less discrete hand-movements (P < 0.001). CONCLUSIONS Mesh reinforcement for incisional reinforcement has a significant impact on the strength of the closure. The utilization of a mesh-application system has the potential to amplify the advantages of mesh reinforcement by providing efficiency and consistency to fixation methods, with similar biomechanical strength to hand-sewn mesh. Additional in vivo analysis and randomized controlled trials are needed to further assess clinical efficacy.
    • Correction