High-throughput testing of urogenital and extragenital specimens for detection of Chlamydia trachomatis and Neisseria gonorrhoeae with cobas® CT/NG

2017 
We compared the analytical and clinical performance of cobas® CT/NG for use on the cobas® 6800/8800 Systems with the cobas® 4800 CT/NG Test from urogenital and extragenital specimens in over 12,000 specimens from both male and female subjects in Germany and the United States. The analytical sensitivity was ≤40 EB/ml for Chlamydia trachomatis (CT) and ≤1 CFU/ml for Neisseria gonorrhoeae (NG). Using clinical specimens, the overall percent agreement with the cobas® 4800 CT/NG Test was >98.5%. Across urogenital specimens, there were 93 discrepant specimens; 76 (93.8%) of 81 CT discrepant specimens were 6800+/4800− and 10 (83.3%) of 12 NG discrepant specimens were 6800+/4800−. Sequencing verified CT results for 45 (61.6%) of 73 samples positive by 6800 and 1 (20%) of 5 positive by 4800. Similarly, 7 (70.0%) of 10 NG samples positive by 6800 and 1 of 2 positive by 4800 were confirmed by sequencing. Among discrepant extragenital specimens (all 6800+/4800−), 7 (50%) of 14 oropharyngeal and 23 (76.7%) of 30 anorec...