Low pO2 Contributes to Potential Error in Oxygen Saturation Calculations Using a Point-of-Care Assay

2018 
The present study addressed the accuracy of calculated oxygen saturation (sO2) using point-of-care (POC) testing compared with measured values on a blood gas analyzer.In total, 3,323 sO2 values were measured in 1,180 patients using a CO-oximeter (ABL 800 Flex; Radiometer, Copenhagen, Denmark). Measured parameters were then used to calculate an expected sO2 for the POC method (Abbott i-STAT; Abbott POC, Princeton, NJ). Cases in which calculated sO2 differed from measured sO2 by 10% or more were analyzed.Of the 3,323 comparisons performed, 260 (8%) showed discrepancies (± ≥10%) between measured and calculated sO2 values. Ninety-four of discrepant measurements (245 of 260) occurred when pO2 was less than 50 mm Hg. pH and bicarbonate distributions shifted to lower values in discrepant vs nondiscrepant cases.Our results suggest that the likelihood of discrepant sO2 is 27% among patients with pO2 less than 50 mm Hg. Direct measurement of sO2 by CO-oximetry is strongly suggested in this clinical scenario.