Initial Stability of Cemented vs Cementless Tibial Components Under Cyclic Load.

2017 
Abstract Background Cement fixation of total knee components remains the gold standard despite resurgence in cementless fixation with the goal of long-term durable fixation. Initial stability is paramount to achieve bony ingrowth of cementless components. Methods Twelve cemented and cementless tibial baseplates were implanted into sawbones and tested using a physiological medial-lateral load distribution for 10,000 cycles to represent 8 weeks of in vivo function. Micromotion was measured at 5 locations around the baseplate during loading. Results Cycling had a significant effect on the change in micromotion between maximum and minimum loads at the anterior, medial, lateral, posteromedial, and posterolateral tray edge locations. A significant effect of fixation technique was detected for the anterior ( P P  = .002), and lateral ( P  = .0056) locations but not for the posteromedial ( P  = .36) or posterolateral ( P  = .82) locations. Differences in micromotion between cemented and cementless components did not exceed 150 μm at any tested location. Conclusion The micromotion experienced by cementless tibial components in the present study may indicate a lower initial mechanical stability than the cemented group. However, this difference in initial stability may be subclinical because the differences between average cemented and cementless micromotion were
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    70
    References
    9
    Citations
    NaN
    KQI
    []