Biomechanical Analysis of Cortical Versus Pedicle Screw Fixation Stability in TLIF, PLIF, and XLIF Applications:

2019 
Study Design:Cadaveric biomechanical study.Objectives:Medial-to-lateral trajectory cortical screws are of clinical interest due to the ability to place them through a less disruptive, medialized exposure compared with conventional pedicle screws. In this study, cortical and pedicle screw trajectory stability was investigated in single-level transforaminal lumbar interbody fusion (TLIF), posterior lumbar interbody fusion (PLIF), and extreme lateral interbody fusion (XLIF) constructs.Methods:Eight lumbar spinal units were used for each interbody/screw trajectory combination. The following constructs were tested: TLIF + unilateral facetectomy (UF) + bilateral pedicle screws (BPS), TLIF + UF + bilateral cortical screws (BCS), PLIF + medial facetectomy (MF) + BPS, PLIF + bilateral facetectomy (BF) + BPS, PLIF + MF + BCS, PLIF + BF + BCS, XLIF + BPS, XLIF + BCS, and XLIF + bilateral laminotomy + BCS. Range of motion (ROM) in flexion-extension, lateral bending, and axial rotation was assessed using pure moments....