Usefulness of a new radiofrequency thermal balloon catheter for pulmonary vein isolation : A new device for treatment of atrial fibrillation

2003 
Introduction: A rapidly firing or triggered ectopic focus located within a pulmonary vein (PV) or close to the PV ostium could induce atrial fibrillation (AF). The aim of this study was to evaluate the efficacy and safety of a radiofrequency thermal balloon catheter for isolation of the PV from the left atrium (LA). Methods and Results: Twenty patients with drug-resistant paroxysmal AF were treated by isolating the superior PVs using an RF thermal balloon catheter. Using a transseptal approach, the balloon, which had an inflated diameter 5 to 10 mm larger than that of the PV ostium, was wedged at the LA-PV junction. It was heated by a very-high-frequency current (13.56 MHZ) applied to the coil electrode inside the balloon for 2 to 3 minutes, and the procedure was repeated up to four times. The balloon center temperature was maintained at 60° to 75°C by regulating generator output. Successful PV isolation was achieved in 19 of the 20 left superior PVs and in all 20 of the right superior PVs and was associated with a decrease in amplitude of the ostial potentials. Total procedure time was1.8 ± 0.5hours, which included22 ± 7minutes of fluoroscopy time. After a follow-up period of8.1 ± 0.8months, 17 patients were free from AF, with 10 not taking any antiarrhythmic drugs and 7 taking the same antiarrhythmic agent as before ablation. Electron beam computed tomography revealed no complications, such as PV stenosis at ablation sites. Conclusion: The PV and its ostial region can be safely and quickly isolated from the LA by circumferential ablation around the PV ostia using a radiofrequency thermal balloon catheter for treatment of AF. (J Cardiovasc Electrophysiol, Vol. 14, pp. 609-615, June 2003)
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    20
    References
    38
    Citations
    NaN
    KQI
    []