Impact of Dual Chamber Pacing on the Incidence of Atrial and Ventricular Tachyarrhythmias in Recipients of Implantable Cardioverter Defibrillators

2005 
Recent observations suggest that frequent dual-chamber pacing in recipients of implantable cardioverter defibrillators (ICD) may adversely influence clinical outcomes. This prospective, multicenter study examined the relationship between the frequency of atrial (%AP) and ventricular pacing (% VP) and the incidence of atrial (AT) and/or ventricular tachyarrhythmias (VT) in a standard ICD population. A total of 141 consecutive patients with primary and secondary ICD indications were studied. Continuous arrhythmia detection with a dual-chamber ICD revealed paroxysmal AT in 60 (43%) and VT in 72 (51%) patients within 6 months of device implantation. Far-field oversensing of ventricular signals occurred in 13% of all "atrial tachy response" mode switches. Without adjustment for covariates, a higher %AP was associated with an increased incidence of AT (P 48% combined with a % VP > 40% was associated with an increased probability for VT. In conclusion, AT and VT occurred frequently within 6 months after dual-chamber ICD implantation. High rates of DDD/R stimulation were associated with a trend toward higher incidence of AT, VT, or both.