オーバーテーブル式 X 線透視装置の表示線量の簡易精度検証方法
2021
The purpose of this study was to establish a simple measurement method to verify the accuracy of incident air kerma (Ka, r) and air kerma area product (PKA) displayed on an over-couch-type X-ray fluoroscopy system. A dosimeter was located at the patient entrance reference point, and the irradiation field size was set to 10×10 cm. A lead plate was placed on the couchtop to protect the image receptor, and the duration of fluoroscopy was set to 1 min. The Ka, r was measured with the proposed method and the Japanese Industrial Standards (JIS) method on three X-ray fluoroscopy units of different manufactures. The effect of backscattered X-rays from the lead plate was calculated using Monte Carlo methods. The errors of the displayed Ka, r and PKA to the measured Ka, r and PKA with our proposed method were calculated. There was no significant difference in the measured Ka, r between the proposed method and the JIS method in all units. The effect of backscattered X-ray was ≤0.5%. The errors of displayed Ka, r and PKA to those measured were in the range of 3.4 to 15.7% and -4.1 to 20.3%, respectively, which met the tolerance for accuracy of ±35% in accordance with the JIS method. We found that our proposed method was simple and that the accuracy of measured values was comparable to that of the JIS method.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
5
References
0
Citations
NaN
KQI