Abstract Surface Plasmon Resonance Biosensors (SPR) are one of the most powerful tools to characterize protein binding, e.g. for drug discovery, like target identification, ligand fishing, assay development, lead selection and manufacturing quality control. However, there is increasing concern about its reproducibility in the light of the reproducibility crisis. Therefore an appropriate analytical instrument qualification (AIQ) is required for quality assurance of SPR instruments. AIQ is a prerequisite for analytical method validation and it is consisting of four parts, Design Qualification (DQ), Installation Qualification (IQ), Operational Qualification (OQ) and Performance Qualification (PQ). PQ regularly executed is supposed to continuously control the performance of the instrument under actual running conditions. In this work a performance qualification method was developed for the SPR instrument Biacore X100. This method is suitable for the routinely control of the instrument performance for antibody-antigen binding measurements. Control charts were designed to get a clearly representable and easy implementable tool to check the critical parameters. These control charts and a straightforward protocol now allow the design and application of an individual performance qualification procedure that can be used in the laboratory routine. They serve as reference for individual standard operation procedures (SOPs).