In this paper, we present a MEMS (micro-electro-mechanical system) implantable blood pressure sensor which has designed and fabricated with consideration of size, design flexibility, and wireless detection. Mechanical and electrical characterizations of the sensor were obtained by mathematical analysis and computer aided simulation. The sensor is composed of two coils and a air gap capacitor formed by separation of the coils. Therefore, the sensor produces its resonant frequency which is changed by external pressure variation. This frequency movement is detected by inductive coupling between the sensor and an external antenna coil. Theoretically analyzed resonant frequency of the sensor under 760 mmHg was calculated to 269.556 MHz. Fused silica was selected as sensor material with consideration of chemical and electrical reaction of human body to the material. 2 mm × 5 mm × 0.5 mm pressure sensors fitted to radial artery were fabricated on the substrates by consecutive microfabrication processes: sputtering, etching, photolithography, direct bonding and laser welding. Resonant frequencies of the fabricated sensors were in the range of 269∼284 MHz under 760 mmHg pressure.