Radiographic assessment of snap-shunt failure: Report of 2 cases

2010 
Accurate assessment of imaging studies in patients with ventriculoperitoneal shunts can be aided by empirical findings. The authors characterize an objective measurement easily performed on head CT scans with the goal of producing clear evidence of shunt fracture or disconnection in patients with a snap shunttype system. The authors describe 2 cases of ventriculoperitoneal shunt failure involving a fracture and a disconnection associated with a snap-shunt assembly. In both cases the initial clinical symptoms were not convincing for shunt malfunction, and the interpretation of the CT finding failed to immediately identify the abnormality. As the clinical picture became more convincing for shunt malfunction, each patient subsequently underwent successful shunt revision. The authors reviewed the CT scans of 10 patients with an intact and functioning snap-shunt system to characterize the normal appearance of the snap-shunt connection. On CT scans the distance between the radiopaque portion of the ventricular...