Validation of GEANT4 simulations for percentage depth dose calculations in heterogeneous media by using small photon beams from the 6-MV Cyberknife: Comparison with photon beam dosimetry with EBT2 film
2015
Percentage depth dose (PDD) distributions in heterogeneous phantoms with lung and soft bone equivalent media are studied by using the GEANT4 Monte Carlo code. For lung equivalent media, Balsa wood is used, and for soft bone equivalent media, a compound material with epoxy resin, hardener and calcium carbonate is used. Polystyrene slabs put together with these materials are used as a heterogeneous phantom. Dose measurements are performed with Gafchromic EBT2 film by using photon beams from the 6-MV CyberKnife at the Seoul Uridul Hospital. The cone sizes of the photon beams are varied from 5 to 10 to 30 mm. When the Balsa wood is inserted in the phantom, the dose measured with EBT2 film is found to be significantly different from the dose without the EBT2 film in and the dose beyond the Balsa wood region, particularly for small field sizes. On the other hand, when the soft bone equivalent material is inserted in the phantom, the discrepancy between the dose measured with EBT2 film and the dose without EBT2 film can be seen only in the region of the soft bone equivalent material. GEANT4 simulations are done with and without the EBT2 film to compare the simulation results with measurements. The GEANT4 simulations including EBT2 film are found to agree well with the measurements for all the cases within an error of 2.2%. The results of the present study show that GEANT4 gives reasonable results for the PDD calculations in heterogeneous media when using photon beams produced by the 6-MV CyberKnife
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
47
References
7
Citations
NaN
KQI