Comparison of the NMIJ and the ARPANSA standards for absorbed dose to water in high-energy photon beams

2015 
: The authors report the results of an indirect comparison of the standards of absorbed dose to water in high-energy photon beams from a clinical linac and (60)Co radiation beam performed between the National Metrology Institute of Japan (NMIJ) and the Australian Radiation Protection and Nuclear Safety Agency (ARPANSA). Three ionisation chambers were calibrated by the NMIJ in April and June 2013 and by the ARPANSA in May 2013. The average ratios of the calibration coefficients for the three ionisation chambers obtained by the NMIJ to those obtained by the ARPANSA were 0.9994, 1.0040 and 1.0045 for 6-, 10- and 15-MV (18 MV at the ARPANSA) high-energy photon beams, respectively. The relative standard uncertainty of the value was 7.2 × 10(-3). The ratio for (60)Co radiation was 0.9986(66), which is consistent with the results published in the key comparison of BIPM.RI(I)-K4.